MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology Recalled by Elekta, Inc. Due to There is a potential that Wedge IDs were...

Date: December 13, 2018
Company: Elekta, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

MOSAIQ¿ Oncology Information System and Sequencer MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed.

Quantity: 3,950 units total (1,747 US)

Why Was This Recalled?

There is a potential that Wedge IDs were not included in the DICOM RT PLAN sent from the TPS, which will cause the field in MOSAIQ to be created with no wedge. Treatment of a field without the planned wedge will receive more dose than the treatment plan indicates.