Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Monaco RTP System. Radiation treatment planning. Recalled by Elekta, Inc. Due to The beam is displayed at the "gantry" angle,...

Date: February 11, 2013
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

Monaco RTP System. Radiation treatment planning.

Quantity: 13

Why Was This Recalled?

The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta, Inc.

Elekta, Inc. has 100 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report