Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Monaco RTP System Recalled by Elekta Inc Due to It is possible that the forced electron density...

Date: September 4, 2019
Company: Elekta Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta Inc directly.

Affected Products

Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Quantity: 49 units

Why Was This Recalled?

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta Inc

Elekta Inc has 10 total recalls tracked by RecallDetector.