Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Elekta MOSAIC Oncology Information System versions: 2.62 Recalled by Elekta, Inc. Due to Malware attack

Date: April 20, 2021
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

Elekta MOSAIC Oncology Information System versions: 2.62, 2.64, 2.70, 2.81, 2.82, and 2.83 hosted on Classic Cloud

Quantity: 40 units

Why Was This Recalled?

Malware attack

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Elekta, Inc.

Elekta, Inc. has 100 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report