Elekta Medical Linear Accelerator (EMLA) Recalled by Elekta, Inc. Due to There is a risk that the Diode D1...

Date: September 17, 2021
Company: Elekta, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Elekta, Inc. directly.

Affected Products

Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD

Quantity: 4,405 units

Why Was This Recalled?

There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

Where Was This Sold?

This product was distributed nationwide across all 50 states.