Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Crimper Model 9100CR26 The Crimper is indicated for use in Recalled by Edwards Lifesciences, LLC Due to Edwards Lifesciences is recalling two lot numbers of...

Date: January 20, 2014
Company: Edwards Lifesciences, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Edwards Lifesciences, LLC directly.

Affected Products

Crimper Model 9100CR26 The Crimper is indicated for use in preparing the Edwards SAPIEN Transcatheter Heart Valve for implantation.

Quantity: 422

Why Was This Recalled?

Edwards Lifesciences is recalling two lot numbers of the 26mm transcatheter heart valve crimper because they may not open sufficiently when the handle is rotated in the full open position. As a result, user may have difficulty fitting a 26mm Sapien transcatheter heart valve into the crimper orifice when preparing the valve for crimping.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Edwards Lifesciences, LLC

Edwards Lifesciences, LLC has 94 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report