Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178 Recalled by Ebi, Llc Due to Broken screw insertor hardness was measured and found...

Date: September 29, 2012
Company: Ebi, Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Ebi, Llc directly.

Affected Products

Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178

Quantity: 15 units

Why Was This Recalled?

Broken screw insertor hardness was measured and found to be below the acceptable limits.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Ebi, Llc

Ebi, Llc has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report