Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment Recalled by EBI, LLC Due to Damage to the lead wire sheath (black cable)...

Date: March 22, 2024
Company: EBI, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact EBI, LLC directly.

Affected Products

Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones. Part Number: 1067718

Quantity: 1412 units ( 2 leads per assembly)

Why Was This Recalled?

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About EBI, LLC

EBI, LLC has 4 total recalls tracked by RecallDetector.