Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Dynavox T15 Speech Generating Device ( including Batteries) Recalled by Dynavox Systems Llc Due to The battery is integral to the device. If...

Date: December 14, 2017
Company: Dynavox Systems Llc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Dynavox Systems Llc directly.

Affected Products

Dynavox T15 Speech Generating Device ( including Batteries)

Quantity: 307

Why Was This Recalled?

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Where Was This Sold?

Nationally

About Dynavox Systems Llc

Dynavox Systems Llc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report