Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) Recalled by Draeger Medical Systems, Inc. Due to Edifier speakers (shipped in or after April 2023)...

Date: January 14, 2025
Company: Draeger Medical Systems, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical Systems, Inc. directly.

Affected Products

Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.

Quantity: 1,013 units

Why Was This Recalled?

Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Draeger Medical Systems, Inc.

Draeger Medical Systems, Inc. has 12 total recalls tracked by RecallDetector.