Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Draeger Jaundice Meter JM-105 The device is intended for use in hospitals Recalled by Draeger Medical Systems, Inc. Due to Users have misinterpreted the display for out of...

Date: May 15, 2018
Company: Draeger Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical Systems, Inc. directly.

Affected Products

Draeger Jaundice Meter JM-105 The device is intended for use in hospitals, clinics or doctor s offices under a physician s supervision / direction to assist clinicians in monitoring of newborn infants. The device is not intended as a standalone for diagnosis of hyperbilirubinemia. It is to be used in conjunction with other clinical signs and laboratory measurements. The Jaundice Meter is not intended for home use. The Jaundice Meter may only be used at the sternum measurement site for Physician s office applications.

Quantity: 6711

Why Was This Recalled?

Users have misinterpreted the display for out of range measurement indicated by the blinking "-0-" to mean a zero measurement

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Draeger Medical Systems, Inc.

Draeger Medical Systems, Inc. has 12 total recalls tracked by RecallDetector.