Draeger Infinity CentralStation (ICS) Recalled by Draeger Medical Systems, Inc. Due to The software on the Infinity CentralStation drops peaks...

Date: April 5, 2023
Company: Draeger Medical Systems, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical Systems, Inc. directly.

Affected Products

Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower

Quantity: 6803 systems

Why Was This Recalled?

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Where Was This Sold?

This product was distributed nationwide across all 50 states.