Draeger Infinity Acute Care System Monitoring Solution Recalled by Draeger Medical Systems, Inc. Due to After a test in which a low SpO2...

Date: December 10, 2013
Company: Draeger Medical Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical Systems, Inc. directly.

Affected Products

Draeger Infinity Acute Care System Monitoring Solution

Quantity: 439

Why Was This Recalled?

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Where Was This Sold?

This product was distributed to 8 states: CA, IL, IN, MA, NH, NY, OK, TX