Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the Recalled by Draeger Medical, Inc. Due to Software 2.51 and Lower with Installed CO2 Measurement...

Date: September 10, 2021
Company: Draeger Medical, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical, Inc. directly.

Affected Products

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Quantity: 6,797 units

Why Was This Recalled?

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Draeger Medical, Inc.

Draeger Medical, Inc. has 78 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report