Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Evita Infinity V500 Ventilator Recalled by Draeger Medical, Inc. Due to The battery capacity of the optional PS500 power...

Date: February 4, 2016
Company: Draeger Medical, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Draeger Medical, Inc. directly.

Affected Products

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Quantity: 2553

Why Was This Recalled?

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Draeger Medical, Inc.

Draeger Medical, Inc. has 78 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report