Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
LFS Flat screen arm system with surgical lamp for Models Aurora series Recalled by DKK Dai-Ichi Shomei Co., Ltd. Due to The stopper block on the LFS bracket could...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact DKK Dai-Ichi Shomei Co., Ltd. directly.
Affected Products
LFS Flat screen arm system with surgical lamp for Models Aurora series, Aurora 2 series, Aurora 3 series, Aurora 4 series, Aurora Astro series, Stellar XL, and Stellar series, LFS Flat Screen Arm System Model Numbers LFSST2323, LFSLFSST29TV23, LFSST29TV23, LFSST23TV, LFSST23, LFSST2323TV, LFSLFSST2323, LDFSLFSLED7TV5, LDFSLFSLED7TV55, LFSLED55, LFSLFSST23, LFSST2923TV, LFSST29TV, LFSST29, LFSST19WC, LFSST2923, LFSLFSST-1, LFSST292323, LFSST1919WC, LFS ST2319WC
Quantity: 256 units
Why Was This Recalled?
The stopper block on the LFS bracket could be broken or missing. Without this stopper block, the flat screen attached to the LFS flat screen system can rotate beyond the original angle limit and it could cause excessive force to be given to the spring shaft of the LFS bracket. Due to the excessive force given to the spring shaft, the spring shaft could break and shear off.
Where Was This Sold?
This product was distributed to 1 state: MI
About DKK Dai-Ichi Shomei Co., Ltd.
DKK Dai-Ichi Shomei Co., Ltd. has 6 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report