Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DiaSorin LIAISON HSV-2 Type Specific IgG Recalled by Diasorin Inc. Due to DiaSorin has identified that lots 136344, 136344A, 136344B,...

Date: December 13, 2023
Company: Diasorin Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Diasorin Inc. directly.

Affected Products

DiaSorin LIAISON HSV-2 Type Specific IgG, REF 310620

Quantity: 31992 units

Why Was This Recalled?

DiaSorin has identified that lots 136344, 136344A, 136344B, 136344C, 136520, 136520A, 136520B, 136520C, 136604, 136703, 136703A, 136703B, 136703C 136775of LIAISON HSV-2 Type Specific IgG assay may produce increased equivocal and/or false positive results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Diasorin Inc.

Diasorin Inc. has 6 total recalls tracked by RecallDetector.

Related Recalls