Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Neoplastine¿ Cl Plus @ (ref.00374) Product Usage: Manual or automated Recalled by Diagnostica Stago, Inc. Due to Confirmed defect of homogeneity where some reagent vials...

Date: October 25, 2017
Company: Diagnostica Stago, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Diagnostica Stago, Inc. directly.

Affected Products

Neoplastine¿ Cl Plus @ (ref.00374) Product Usage: Manual or automated determination of the prothrombin time (PT).

Quantity: 1636 Kits

Why Was This Recalled?

Confirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).

Where Was This Sold?

Nationally

About Diagnostica Stago, Inc.

Diagnostica Stago, Inc. has 25 total recalls tracked by RecallDetector.