Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Recalled by Dexcom, Inc. Due to A software defect in version v1.15.0 of the...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Dexcom, Inc. directly.
Affected Products
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
Quantity: 333
Why Was This Recalled?
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).
Where Was This Sold?
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.
About Dexcom, Inc.
Dexcom, Inc. has 14 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report