Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeVilbiss Recalled by DeVilbiss Healthcare LLC Due to The firm has received complaints of the device...

Date: November 18, 2022
Company: DeVilbiss Healthcare LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DeVilbiss Healthcare LLC directly.

Affected Products

DeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q

Quantity: 1004315 (US); 414559 (OUS)

Why Was This Recalled?

The firm has received complaints of the device becoming hot. The firm has updated the user manual to include warnings that applied parts in contact with the patient may in some situations exceed 41 degrees Celsius (105.8 degrees Fahrenheit), clarification on expected service life, and the need for maintenance to certain components.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DeVilbiss Healthcare LLC

DeVilbiss Healthcare LLC has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report