Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeRoyal Tonsillectomy Tray Recalled by DeRoyal Industries Inc Due to DeRoyal is recalling certain lots of surgical procedure...

Date: November 28, 2022
Company: DeRoyal Industries Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DeRoyal Industries Inc directly.

Affected Products

DeRoyal Tonsillectomy Tray, REF 89-10698.01

Quantity: 320 units

Why Was This Recalled?

DeRoyal is recalling certain lots of surgical procedure pack products that contain a ConMed Foot Controlled Electrosurgical Suction Coagulator. The recall is due to ConMed receiving reports that the suction ports may be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use.

Where Was This Sold?

This product was distributed to 2 states: SC, VA

Affected (2 states)Not affected

About DeRoyal Industries Inc

DeRoyal Industries Inc has 225 total recalls tracked by RecallDetector.