Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeRoyal SafeLiner Suction Canister Recalled by DeRoyal Industries Inc Due to The reason for the voluntary recall is due...

Date: February 9, 2023
Company: DeRoyal Industries Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DeRoyal Industries Inc directly.

Affected Products

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Quantity: 27,200 devices

Why Was This Recalled?

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DeRoyal Industries Inc

DeRoyal Industries Inc has 225 total recalls tracked by RecallDetector.

Related Recalls