Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeRoyal limb holders: (1) DeRoyal Limb Holder Recalled by DeRoyal Industries Inc Due to The webbing strap was laced improperly through the...

Date: July 15, 2022
Company: DeRoyal Industries Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DeRoyal Industries Inc directly.

Affected Products

DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.

Quantity: 36 eaches and 7 pairs

Why Was This Recalled?

The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DeRoyal Industries Inc

DeRoyal Industries Inc has 225 total recalls tracked by RecallDetector.