Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit Recalled by DeRoyal Industries Inc Due to The unit displays an error code of hose...

Date: September 17, 2013
Company: DeRoyal Industries Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DeRoyal Industries Inc directly.

Affected Products

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.

Quantity: 2 units

Why Was This Recalled?

The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DeRoyal Industries Inc

DeRoyal Industries Inc has 225 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report