Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

VELYS Robotic-Assisted Solution Base Product No.: 451570100 Recalled by DePuy Orthopaedics, Inc. Due to System software v1.5.1 has a system software issue...

Date: March 11, 2022
Company: DePuy Orthopaedics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DePuy Orthopaedics, Inc. directly.

Affected Products

VELYS Robotic-Assisted Solution Base Product No.: 451570100

Quantity: US: 68 units; OUS: 10 units

Why Was This Recalled?

System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DePuy Orthopaedics, Inc.

DePuy Orthopaedics, Inc. has 154 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report