Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

REF 623401 L S-ROM Noiles Rotating Hinge Recalled by DePuy Orthopaedics, Inc. Due to DePuy Orthopaedics has identified the potential for holes...

Date: February 24, 2014
Company: DePuy Orthopaedics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DePuy Orthopaedics, Inc. directly.

Affected Products

REF 623401 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time.

Quantity: 129

Why Was This Recalled?

DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM¿ Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DePuy Orthopaedics, Inc.

DePuy Orthopaedics, Inc. has 154 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report