Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LCS Complete Knee Revision System VVC Insert Product Usage: LCS Recalled by DePuy Orthopaedics, Inc. Due to DePuy Orthopaedics, Inc. is voluntarily recalling one lot...

Date: November 8, 2016
Company: DePuy Orthopaedics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DePuy Orthopaedics, Inc. directly.

Affected Products

LCS Complete Knee Revision System VVC Insert Product Usage: LCS VVC Insert is used in Knee Revision Surgery

Quantity: 1

Why Was This Recalled?

DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which prevents the insert from properly seating in the mating tibial tray.

Where Was This Sold?

This product was distributed to 1 state: TN

Affected (1 state)Not affected

About DePuy Orthopaedics, Inc.

DePuy Orthopaedics, Inc. has 154 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report