Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ATTUNE Intuition Distal Femoral Jig Recalled by DePuy Orthopaedics, Inc. Due to Issuing a device correction because if the pin...

Date: October 15, 2014
Company: DePuy Orthopaedics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DePuy Orthopaedics, Inc. directly.

Affected Products

ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyethylene bag

Quantity: 3,082

Why Was This Recalled?

Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About DePuy Orthopaedics, Inc.

DePuy Orthopaedics, Inc. has 154 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report