Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DELFI SINGLE USE VARIFIT TOURNIQUET CUFF SINGLE PORT SINGLE BLADDER Recalled by Delfi Medical Innovations, Inc. Due to Delfi Medical Innovations Inc has discovered that tourniquets...

Date: April 3, 2019
Company: Delfi Medical Innovations, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Delfi Medical Innovations, Inc. directly.

Affected Products

DELFI SINGLE USE VARIFIT TOURNIQUET CUFF SINGLE PORT SINGLE BLADDER CONTOUR THIGH CUFF WITH MATHCING LIMP PROTECTION SLEEVE Rx Only, Sterile

Quantity: 24 units

Why Was This Recalled?

Delfi Medical Innovations Inc has discovered that tourniquets cuffs in certain lots of the VariFit Sterile Single Use Thigh Cuffs, Single Port, Single Bladder models may have been manufactured with an inadequate perimeter weld on the cuff. Once applied to a patients thigh and inflated to a pressure to occlude blood flow in the thigh distal to the cuff, after a period of time the cuff may leak to the extent that the tourniquet instrument supplying air pressure to the cuff may not be able to adequately control and regulate pressure in the cuff. This may result in some blood flow in the thigh distal to the cuff.

Where Was This Sold?

US: Ohio OUS: Austrailia

About Delfi Medical Innovations, Inc.

Delfi Medical Innovations, Inc. has 2 total recalls tracked by RecallDetector.

Related Recalls