Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Cardinal Health Jackson-Pratt Channel Drain 15FR Recalled by Degania Silicone, Ltd. Due to 15FR and 19FR round fluted drain channel drain...

Date: June 28, 2019
Company: Degania Silicone, Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Degania Silicone, Ltd. directly.

Affected Products

Cardinal Health Jackson-Pratt Channel Drain 15FR, Full flutes, with trocar Product Number: JP-2189 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Quantity: 6080

Why Was This Recalled?

15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Degania Silicone, Ltd.

Degania Silicone, Ltd. has 43 total recalls tracked by RecallDetector.