Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

DDU-2450 Lifeline ECG Model Numbers DDU-A2450EN DDU-E2450DE DDU-E2450EN DDU-E2450FR DDU-E2450TH Recalled by Defibtech, LLC Due to A component issue may cause the AED to...

Date: April 27, 2020
Company: Defibtech, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Defibtech, LLC directly.

Affected Products

DDU-2450 Lifeline ECG Model Numbers DDU-A2450EN DDU-E2450DE DDU-E2450EN DDU-E2450FR DDU-E2450TH Cat. No. DCF-A2460EN DCF-A2463EN DCF-E2460DE DCF-E2460EN DCF-E2460FR DCF-E2460TH UDI 00815098020287, 00815098020317 Automated External Defibrillators (AED) are indicated for use on victims of sudden cardiac arrest (SCA) who are: - unconscious and unresponsive, not breathing or not breathing normally.

Quantity: 860

Why Was This Recalled?

A component issue may cause the AED to abort a shock delivery or reset unexpectedly, which may cause an affected device to fail to delivery shock, and/or to fail to deliver shock in range.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Defibtech, LLC

Defibtech, LLC has 14 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report