Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

GE Healthcare Giraffe Omnibed Carestation CS1 Recalled by DATEX--OHMEDA, INC. Due to Prior to current established manufacturing practices, higher levels...

Date: June 6, 2023
Company: DATEX--OHMEDA, INC.
Status: Completed
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact DATEX--OHMEDA, INC. directly.

Affected Products

GE Healthcare Giraffe Omnibed Carestation CS1, combination infant incubator and infant warmer. Model Numbers: 1) 2082844-001-01040418; 2) 2082844-001-01046784; 3) 2082844-001-01052324; 4) 2082844-001-01052572; 5) 2082844-001-01055337; 6) 2082844-001-01065693; 7) 2082844-001-01069008; 8) 2082844-001-01073750; 9) 2082844-001-01086553; 10) 2082844-001-01089393; 11) 2082844-001-01090414; 12) 2082844-001-01092632; 13) 2082844-001-01111482; 14) 2082844-001-01115741; 15) 2082844-001-01122391; 16) 2082844-001-01128579; 17) 2082844-001-01128725; 18) 2082844-001-01131573; 19) 2082844-001-01140642; 20) 2082844-001-01165870; 21) 2082844-001-100397; 22) 2082844-001-100676; 23) 2082844-001-100740; 24) 2082844-001-100743; 25) 2082844-001-100744; 26) 2082844-001-101075; 27) 2082844-001-101149; 28) 2082844-001-101199; 29) 2082844-001-101202; 30) 2082844-001-101203; 31) 2082844-001-101356; 32) 2082844-001-101366; 33) 2082844-001-101656; 34) 2082844-001-102136; 35) 2082844-001-102589; 36) 2082844-001-102858; 37) 2082844-001-526493; 38) 2082844-001-533346; 39) 2082844-001-570904; 40) 2082844-001-615985; 41) 2082844-001-616544; 42) 2082844-001-674814; 43) 2082844-001-701655; 44) 2082844-001-704407; 45) 2082844-001-709042; 46) 2082844-001-857748; 47) 2082844-001-888278; 48) 2082844-001-941288; 49) 2082844-001-965539; 50) 2082844-001-978713; 51) 2082844-001-982099; 52) 2082844-001-982310; 53) 2082844-001-982985; 54) 2082844-001-982988; 55) 2082844-001-983509; 56) 2082844-001-983643; 57) 2082844-001-983679; 58) 2082844-001-984096; 59) 2082844-001-984948; 60) 2082844-001-987480; 61) 2082844-001-987855; 62) 2082844-001-993967; 63) 2082844-001-994441; 64) 2082844-001-995782; 65) 2082844-001-998377; Radiant infant warmer

Quantity: 769 units

Why Was This Recalled?

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

Where Was This Sold?

worldwide

About DATEX--OHMEDA, INC.

DATEX--OHMEDA, INC. has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report