Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Cardiosave Li-Ion Battery Pack Recalled by Datascope Corp. Due to A Cardiosave Lithium-Ion Battery Pack used during protocol...

Date: September 21, 2021
Company: Datascope Corp.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Datascope Corp. directly.

Affected Products

Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55. 2. Cardiosave Rescue. Model Number: 0998-00-0800-83.

Quantity: 131 batteries, 11,517 pumps

Why Was This Recalled?

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Datascope Corp.

Datascope Corp. has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report