Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31 Recalled by Datascope Corp. Due to Firm is initiating a correction due to the...

Date: November 15, 2021
Company: Datascope Corp.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Datascope Corp. directly.

Affected Products

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Quantity: 5,566 (US); 6,354 (OUS)

Why Was This Recalled?

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Datascope Corp.

Datascope Corp. has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report