Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) Recalled by Datascope Corp. Due to Battery short-run times and unanticipated stoppage may be...

Date: June 17, 2019
Company: Datascope Corp.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Datascope Corp. directly.

Affected Products

Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53 For cardiovascular use.

Quantity: 6,197

Why Was This Recalled?

Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed. Inadequate labeling is being targeted for this recall, and design updates will be handled in a subsequent recall.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Datascope Corp.

Datascope Corp. has 66 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report