Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

PT Link Waterbath Recalled by Dako North America Inc. Due to Dako is recalling the PT Link Waterbath because...

Date: May 1, 2008
Company: Dako North America Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Dako North America Inc. directly.

Affected Products

PT Link Waterbath, Model#PT100. Lab Vision PT Module Plus provides an automated standarized alternative to labor intensive and time consuming dewaxing and epitote retrieval procedures.

Quantity: 136 units

Why Was This Recalled?

Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Dako North America Inc.

Dako North America Inc. has 8 total recalls tracked by RecallDetector.