ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 Recalled by Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.) Due to Due to interference with the Live Listen feature...

Date: April 21, 2022
Company: Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.)
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.) directly.

Affected Products

ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 or earlier

Quantity: 41 devices (out of 1,260 users)

Why Was This Recalled?

Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.

Where Was This Sold?