Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions Recalled by Cytocell Ltd. Due to individual components have been labelled with incorrect colours....

Date: January 18, 2022
Company: Cytocell Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cytocell Ltd. directly.

Affected Products

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Quantity: 19 kits

Why Was This Recalled?

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Cytocell Ltd.

Cytocell Ltd. has 14 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report