Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.
CMS-4873 Custom CT Biopsy Tray Recalled by Custom Medical Specialties, Inc. Due to The firm saw on the FDA web site...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Custom Medical Specialties, Inc. directly.
Affected Products
CMS-4873 Custom CT Biopsy Tray, CMS-4873-R1 Custom CT Biopsy Tray containing 4 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep
Quantity: 538 cases
Why Was This Recalled?
The firm saw on the FDA web site a recall by the manufacturer of the Povidone Iodine Prep Solution due to no microbial testing and product did not meet proper finished goods specifications.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Custom Medical Specialties, Inc.
Custom Medical Specialties, Inc. has 21 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report