Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Sensica Urine Output System Recalled by C.R. Bard Inc Due to BD (C.R. Bard) has received user complaints reporting...

Date: November 15, 2023
Company: C.R. Bard Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact C.R. Bard Inc directly.

Affected Products

Sensica Urine Output System, Catalog Number SCCS1002

Quantity: 148 units

Why Was This Recalled?

BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About C.R. Bard Inc

C.R. Bard Inc has 65 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report