Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Arctic Sun Temperature Management System Recalled by C.R. Bard, Inc. Due to The product may contain electronic components that lead...

Date: November 9, 2015
Company: C.R. Bard, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact C.R. Bard, Inc. directly.

Affected Products

Arctic Sun Temperature Management System, 5000, 100-120V and 220-240V The product is a thermal regulating system indicated for monitoring and controlling patient temperature

Quantity: 2613

Why Was This Recalled?

The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell battery drainage could render the device unresponsive upon System startup.

Where Was This Sold?

This product was distributed to 47 states: AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, DC

Affected (47 states)Not affected

About C.R. Bard, Inc.

C.R. Bard, Inc. has 51 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report