Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

McGrath Mac Video Laryngoscope Recalled by Covidien Due to Battery management system within Laryngoscope devices may deplete...

Date: July 9, 2024
Company: Covidien
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Covidien directly.

Affected Products

McGrath Mac Video Laryngoscope, REF: 300-000-000 , Non-Sterile, Rx Only

Quantity: 75,544 units

Why Was This Recalled?

Battery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Covidien

Covidien has 21 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report