Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tri-Staple 2.0 Black Intelligent Reload Recalled by Covidien, LP Due to Affected lots have the potential for a broken...

Date: January 25, 2023
Company: Covidien, LP
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Covidien, LP directly.

Affected Products

Tri-Staple 2.0 Black Intelligent Reload, Product Number SIG60AXT

Quantity: 264 US; 5056 OUS

Why Was This Recalled?

Affected lots have the potential for a broken sled vane, which may cause the reload to misfire leading to non-functional staple line closure, transecting tissue without forming staples, and tissue hang-up. These conditions may be associated to a delay to treatment, unspecified infection, hemorrhage/blood loss/bleeding, failure to anastomose, peritonitis, sepsis, pneumothorax, tissue trauma, or death.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Covidien, LP

Covidien, LP has 68 total recalls tracked by RecallDetector.

Related Recalls