Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Medi-Trace Cadence Pediatric Radiotransparent Defibrillation Electrode Product ID: 22550P. Intended Recalled by Covidien LLC Due to Defibrillation Electrodes may arc/spark result in thermal damage...

Date: April 8, 2013
Company: Covidien LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Covidien LLC directly.

Affected Products

Medi-Trace Cadence Pediatric Radiotransparent Defibrillation Electrode Product ID: 22550P. Intended for use in cardioversion, pacing, ECG monitoring, and defibrillation procedures.

Quantity: 7485 (sets of 2)

Why Was This Recalled?

Defibrillation Electrodes may arc/spark result in thermal damage to the leadwire, which could render the device incapable of delivering the appropriate energy or shock to the patient

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Covidien LLC

Covidien LLC has 248 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report