Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Recalled by Covidien Llc Due to Intermittent risk of electric shock to users if...

Date: November 12, 2021
Company: Covidien Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Covidien Llc directly.

Affected Products

Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA

Quantity: 2 units OUS

Why Was This Recalled?

Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit

Where Was This Sold?

Brazil

About Covidien Llc

Covidien Llc has 59 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report