Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MINC+ Benchtop Incubator Recalled by Cook Medical Llc Due to The MINC+ device is susceptible to losing temperature...

Date: December 7, 2023
Company: Cook Medical Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cook Medical Llc directly.

Affected Products

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Quantity: 14 US; 101 OUS

Why Was This Recalled?

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Where Was This Sold?

This product was distributed to 5 states: CA, IN, LA, MD, NY

Affected (5 states)Not affected

About Cook Medical Llc

Cook Medical Llc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report