Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Tornado Embolization Microcoil Recalled by Cook Incorporated Due to The loading cartridges included with these devices may...

Date: May 5, 2022
Company: Cook Incorporated
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cook Incorporated directly.

Affected Products

Tornado Embolization Microcoil, RPN GPN MWCE-18S-5/2-TORNADO G08356 MWCE-18S-6/2-TORNADO G08259 MWCE-18S-3/2-TORNADO-01 G47416 MWCE-18S-4/2-TORNADO-01 G47417 MWCE-18S-3/2-TORNADO-081800 G13102 MWCE-18S-3/2-TORNADO G08261 MWCE-18S-4/2-TORNADO-081800 G13103 MWCE-18S-4/2-TORNADO G08357 MWCE-18S-5/2-TORNADO-01 G47418 MWCE-18S-6/2-TORNADO-01 G47419 MWCE-18S-6/2-TORNADO-081800 G13105 MWCE-18S-5/2-TORNADO-081800 G13104 MWCE-18S-2/4-TORNADO-LEF G09945 MWCE-18S-2/5-TORNADO-LEF-081800 G13108 MWCE-18S-4/3-TORNADO G12924 MWCE-18S-2/3-TORNADO-LEF G09944 MWCE-18S-2/3-TORNADO-LEF-081800 G13106 MWCE-18S-2/4-TORNADO-LEF-081800 G13107

Quantity: 106,033

Why Was This Recalled?

The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Cook Incorporated

Cook Incorporated has 64 total recalls tracked by RecallDetector.