Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Vue Optic Visualization Source Recalled by Cook Inc. Due to Cook Inc. is initiating a voluntary recall of...

Date: May 3, 2017
Company: Cook Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cook Inc. directly.

Affected Products

Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages

Quantity: 1328 units

Why Was This Recalled?

Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.

Where Was This Sold?

This product was distributed to 7 states: FL, IN, LA, MS, NC, TX, VA

Affected (7 states)Not affected

About Cook Inc.

Cook Inc. has 262 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report