Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Fixed Core Straight Safety Wire Guide Recalled by Cook Inc. Due to The wire guide may be incorrectly loaded into...

Date: January 4, 2019
Company: Cook Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Cook Inc. directly.

Affected Products

Fixed Core Straight Safety Wire Guide, Guidewire, Global Product Number G14285

Quantity: 15

Why Was This Recalled?

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected