Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MINI INFANT < 3KG Recalled by ConMed Corporation Due to Wire sets of certain devices were assembled with...

Date: February 23, 2018
Company: ConMed Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact ConMed Corporation directly.

Affected Products

MINI INFANT < 3KG, RADIOTRANSLUCENT ELECTRODE, Universal/Anderson Connector, REF/Catalog Number 2602 Indicated for use in external pacing, defibrillation and monitoring applications as a non-sterile, disposable device for single pediatric use, only. The PadPro 2602 electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (non-invasive) cardiac pacemaker and the pediatric patients skin. The electrode is intended for use on pediatric infants whose weight is less than 10 kg. (22 lbs.).

Quantity: 18,720 units in total

Why Was This Recalled?

Wire sets of certain devices were assembled with adhesive on the contact wires, which may impede conductivity and could cause failure

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About ConMed Corporation

ConMed Corporation has 69 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report